Mostrar el registro sencillo del ítem

dc.contributores-ES
dc.creatorMainegra, Ernesto
dc.creatorCapote, Roberto
dc.date2016-11-17
dc.identifierhttps://revistas.tec.ac.cr/index.php/tec_marcha/article/view/2858
dc.descriptionThe therapy beam from the 60Co unit used at INOR was modeled using the EGS Monte Carlo radiation transport system. The rectangular geometry of the collimating system is fully simulated and the effect of considering the outer collimator a set of stepped collimators rather than a collimator with a smooth surface was estimated. Variance reduction techniques are widely used to increase the efficiency of our calculations. Even for the dose at very shallow depths (<0.5 mm) the agreement of our calculations with the experimental results is very good. The shift of peak dose to about 1 mm due to electron contamination is observed.es-ES
dc.formatapplication/pdf
dc.languagespa
dc.publisherEditorial Tecnológica de Costa Ricaes-ES
dc.relationhttps://revistas.tec.ac.cr/index.php/tec_marcha/article/view/2858/2620
dc.rightsCopyright (c) 2016 Revista Tecnología en Marchaes-ES
dc.sourceRevista Tecnología en Marcha; Vol 13 edición especial 1 (1997); Pág. 134-139es-ES
dc.source2215-3241
dc.source0379-3982
dc.subjectes-ES
dc.subjectes-ES
dc.titleRealistic Monte Carlo simulation of the INOR Therapy Unit and electron contamination calculation for a board 60Co Beames-ES